In recent years, the number of medical devices designed to be used  within a home environment has increased substantially; this is why it has been necessary the introduction of a specific standards, IEC 60601-1-11, that regulates the safety of healthcare products used at home. These types of devices are normally used  in a uncontrolled environment by a lay user, often without  the help of any healthcare professional and without prior specific training to the end users. 

We have already been talking about design validation and the general standard IEC 60601-1, the defines safety and general performance for the electro medical devices, and more in general, of everything is related to the design control process, from design plan to design review.

Obviously the home environment is really different from the clinical environment and devices intended to be used  at home have to overcome a series of challenges related to the safety of the device. The presence of children and pets may have an impact on device operations; moreover the presence of dust and possibility that the devices come  in contact with water or other liquid need as well to be considered as at home  we  do not have all the controls that a healthcare setting may have in place. 

In this article we will  go  through the main requirements associated to IEC 60601-1-11, discussing some of the specific requested from the standard and the tests that need to be performed to demonstrate compliance.

Environmental conditions for home devices according to IEC 60601-1-11

In the  IEC 60601-1-11, particular attention is given to transport and  storage conditions of the device. The standard specifies that, unless the device can be considered  stationary, the transport and storage conditions have to be aligned to the following requirements: 

  • – 25°C to +5°C, and
  • –  + 5 °C to + 35 °C at a relative humidity up to 90 %, non-condensing;
  • –  >35°C to 70°C at a water  vapour pressure up to 50 hPa

These conditions applies after having removed the device from its initial packaging. 

The  standard accepts the use of a  more restricted range, that however needs to be justified in the risk management file and properly included in the labelling material of the device. 

Similar  requirements are present in the standard in relation to environmental operating conditions. 

Requirements for Supply Mains for Medical  equipments used in home environments are also specified in the IEC 60601-1-11 standard, specifically in the section 4.1. The 60601-1 general standard requires, for any general active medical devices, supply mains to have a 10 percent tolerance from nominal. However, mains voltages in the home may be less predictable. For this reason, the 60601-1-11 collateral standard assumes nominal mains voltages may exhibit a tolerance of +10 percent/-15 percent, corresponding to an operational input voltage range of 85 to 264 VAC for universal line input devices.

Classification  of Medical Equipment according to IEC 60601-1-11

In the chapter 6 of the standard, the requirements associated to the classification of devices to be used within a home environments are reported.

Specifically power supplies for medical device equipments can be Class I or Class II. Class I power supplies protect users via a single layer of basic insulation and a protective Earth ground connection, which keeps users safe by providing a low impedance path to zero potential for fault currents to follow. At the same time, class II power supplies utilize additional basic insulation or reinforced insulation to ensure that with or without a reliable Earth ground connection, users are properly protected from hazardous voltages.

The standard requires home devices to be Class II or internally powered; this basically means that the standard does not allow OEMs to rely on in-home earth group connection, that are often unreliable expecially in older homes.

The IEC 60601-1-11 defines as well the level of ingress protection related to home medical devices, specifically in the section 8.3 of the standard with the chapter entitled Additional requirements for ingress of water or particulate matter into ME EQUIPMENT and ME SYSTEMS. Specifically, the requirement is to have at least a level of ingress protection equal to 22.

In general, medical device enclosures need to be classified according to their requirements with the classification per IEC 60529, or IP rating (ingress protection rating). The IP ratings consist of two numbers: the first, from 0 to 6, deals with the degree of particulate ingress, the second, from 0 to 8, deals with liquid ingress. Both start at a rating of 0, which is unprotected, and go up from there.

Subscribe to 4EasyReg Newsletter

4EasyReg is an online platform dedicated to Quality & Regulatory matters within the medical device industry. Have a look to all the services that we provide: we are very transparent in the pricing associated to these consulting services.

Within our WebShop, a wide range of procedures, templates, checklists are available, all of them focused on regulatory topics for medical device compliance to applicable regulations. Within the webshop, a dedicated section related to cybersecurity and compliance to ISO 27001 for medical device organizations is also present.

As one of the leading online platforms in the medical device sector, 4EasyReg offers extensive support for regulatory compliance. Our services cover a wide range of topics, from EU MDR & IVDR to ISO 13485, encompassing risk management, biocompatibility, usability, software verification and validation, and assistance in preparing technical documentation for MDR compliance.

Do not hesitate to subscribe to our Newsletter!

You have successfully subscribed to the newsletter

There was an error while trying to send your request. Please try again.

4EasyReg will use the information you provide on this form to be in touch with you and to provide updates and marketing.